Service Program
3
Perfusor® compact, 3.0 gb 3- 5
-
Syringes
➨
Syringe selection
-
Constants
➨
Service interval
4. Menu item
EEPROM
➨
Write
transmits data to the device.
Menu item
File
➨
Save
saves the data to the hard disk.
5. Enter the user number 0 upon query.
6. Check unit according to the procedure instructions for
inspection (see “Procedural Instructions for Inspection after
Modifications via the Service Program” ➨ p. 3 - 13).
Unit Calibration
The unit must be calibrated (see “Unit Calibration” ➨ p. 3 - 16)
after the E-Module or the drive has been replaced or the bolus
rate changed.
Defaut Data
The Service Program contains the Default.dat file with the factory
settings of the unit. These values can be adjusted via the Syringe
or Modes menu if required.
Max. delivery rate (basal rate) . . . . . . . . . . . . . . . . . . . . 99.9 ml/h
Bolus rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800 ml/h
Staff call [Version 5.5004]. . . . . . . . . . dynamic without off alarm
[Version 2.1.0.1] . . . . . . . . . . . . . . . . dynamic with pre-alarm
Alarm tone[Version 5.5004] . . . . . . . . . . . . 0 (3 Hz interval tone)
Tone when alarms output [Version 2.1.0.1] . . . . . . . . . . . . . . .3 Hz
Tone when pre-alarms output [Version 2.1.0.1] . . . . . . . . . . static
Pressure stage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Dianet address [Version 5.5004] . . . . . . . . . . . . . . . . . . . . . . . . . 1
Syringe selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table
Service interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20440 hrs.
What To Do If ... (Trouble Shooting)
... the length calibration does not start?
Could communication be started successfully? Does the motor
still not start?
Then: Select Termination. Switch off pump. Repeat
communication start. Switch pump on again.
3.0