Perfusor® compact, 3.0 gb 3- 1
3
3-Service Program
Software Compatibility
Designation Ord. No.
Interface cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 1661
Compatibility
Unit Software Service Program Ord. - No.
PLAA00035.5 5.3502 . . . . . . . . . . . On request
PLAA00040.0 5.5004
†
. . . . . . . . . . 3452 0474
PLAA00041.0 5.5004
†
. . . . . . . . . . 3452 0474
PLAA00063.3 2.1.0.1 . . . . . . . . . . . 3452 0474
PLAA00070.0 (Dianet
Star
) 2.1.0.1 . . . . . . . . . . . 3452 0474
PLAA00071.1 (Dianet
Star
) 2.1.0.1 . . . . . . . . . . . 3452 0474
PLAA00072.1 (Dianet
Star
) 2.1.0.1 . . . . . . . . . . . 3452 0474
PLAA01070 (new hardware) 2.1.0.1 . . . . . . . . . . . 3452 0474
Note
The following description refers to the versions 5.5004 and
2.1.0.1. of the Service Program.
The relevant version is given in square brackets in the illustrations
of each user interface and where the description contains
differences.
Introduction
The Service Program runs in a user-friendly way on a PC. All
functions are easy to operate in the pulldown-menus as in
Windows.
WARNING
!
Service mode is only authorized for service tasks.
DANGER OF INJURY IF USED ON THE PATIENT!
- Never connect the MFC service connector or the service cable
when a patient is connected to the unit!
- First switch the unit off before any further use after working
with the MFC service connector.
- Carry out a check according to the procedural instructions for
inspection after the Service Program was run. (see
“Procedural Instructions for Inspection after Modifications
via the Service Program” ➨ p. 3 - 13).
†
The previous version 5.5004 is included in the installation package for 2.1.0.1.
3 - 18
3.0