Software
2
2- 4 Perfusor® compact, 2.1 gb
Device alarms of the control microprocessor
LC display Description
128 Unexpected reset
129 Unexpected hardware interrupt
130 Access of zero pointer
131 Attempted division by zero
132 Internal software error
134 State/motor state
135 Invalid variable values
136 Invalid operating condition
137 Illegal mode – port value
138 H8 indicates GA F14_H8GA_K16
150 Different software versions
151 Double CRC error
153 Different states
154 Different rates
155 Different F-mode
156 Different mode values
157 Different alarm recognition
158 Different alarm clearance
159 Err. current volume
160 Err. preselected volume
161 Err. volume per step
170 Sensor sync. failed
171 ... 174 Sensor - dark test error
180 ROM test error
181 RAM test error
182 Keyboard test error column
183 Dynamic memory test
184 Motor test no sync
185 Keyboard test error
186 Timer test error
187 CPU test error
188 Battery test error
191 Different software buttons
Table 2 - 2 (Part 1 of 2)
2.1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun perfusor compact and is the answer not in the manual?

Braun perfusor compact Specifications

General IconGeneral
ModelPerfusor compact
Occlusion PressureUp to 1200 mmHg
AlarmsOcclusion
Syringe CompatibilityVarious sizes from 5 ml to 50 ml