Service Program
3
Perfusor® compact, 3.0 gb 3- 9
Calibration Menu
(see “Unit Calibration” ➨ p. 3 - 16)
1.
Serial number
(F4)
Enter the serial number when the E-Module is exchanged as
otherwise the EEPROM cannot be written (Dianet type =
1200).
2.
Length calibration
(F5)
The position of the prealarm light barriers and the drive end
is determined by length calibration. The motor steps
determined are displayed after calibration is terminated.
3.
Pressure calibration
(F6)
The motor parameters for setting the 3 pressure stages and
the correct switch-off in bolus mode is determined by
pressure calibration.
4.
Overload check
(see “Overload Check” ➨ p. 3 - 17)
The dynamic pressure test is used to determine whether the
unit was damaged after having been dropped, due to a shock
or impact or when the drive head was dismounted. The drive
must build-up a pressure of ≥1.6 bar, and the positive locking
sensor must not open.
5.
Parameters
Displays the calibration parameters.
Service Menu
1.
Service Values
or
Processing values
The service values are displayed. These values cannot be
changed. When the default data was specified the service
values are set to zero.
The last max. 20 service alarms and max. 10 device alarms are
logged in the alarm tables.
Fig.: 3 - 7
[Version 5.5004] [Version 2.1.0.1]
Fig.: 3 - 8
[Version 5.5004] [Version 2.1.0.1]
3.0