4
4- 4 Perfusor® compact, 2.1 gb
Close Unit
1. Close unit in reverse order of opening.
Note
Do not squeeze motor cable.
Checks after repair
Procedural instructions (see "Procedural Instructions for
Inspection after Modifications via the Service Program"
p. 3 - 13).
A calibration in the Service Program is to be carried out if a new
E-Module is installed or the drive is replaced (see "Service
Program" p. 3 - 1).