51 Alarms
Clearway 2 User Manual
DOC-001000 Ver 6
Incorrect Circuit in Use
No exhalation valve is detected in the breathing
circuit
Exhalation valve not fitted.
Incorrect breathing circuit being used.
Treatment continues. Alarm continues until
condition resolved or treatment ends
“Incorrect Circuit in use!
- Ensure IPPB valve is fitted”
Internal Over Temperature
Internal temperature is > 60
º
Ambient temperature is high.
Device is positioned in sunlight or close to a
heat source.
Air inlet filters are blocked (e.g. placed on a
bed or soft surface)
Switch device OFF. Leave for 30 minutes”