51 Alarms
Clearway 2 User Manual
DOC-001000 Ver 6
Incorrect Circuit in Use
Item
Description
Definition
No exhalation valve is detected in the breathing
circuit
Priority
Medium
Modes
IPPB
Possible Cause
Exhalation valve not fitted.
Incorrect breathing circuit being used.
Clearway 2 action
Treatment continues. Alarm continues until
condition resolved or treatment ends
On Screen Message
“Incorrect Circuit in use!
- Ensure IPPB valve is fitted”
Internal Over Temperature
Item
Description
Definition
Internal temperature is > 60
º
C.
Priority
Low
Modes
All
Possible Cause
Ambient temperature is high.
Device is positioned in sunlight or close to a
heat source.
Air inlet filters are blocked (e.g. placed on a
bed or soft surface)
Clearway 2 action
Treatment will continue
On Screen Message
“Over Temperature!
Switch device OFF. Leave for 30 minutes”