Alarms 54
DOC-001000 Ver 6 Clearway 2 User Manual .
NIV Treatment Ending Soon
Item
Description
Definition
NIV treatment is about to end
Priority
Warning
Modes
NIV
Possible Cause
NIV mode is ending soon
Clearway 2 action
Treatment continues and message will disappear
after 5 seconds.
On Screen Message
“NIV Treatment is ending soon”
NIV Treatment has Ended
Item
Description
Definition
NIV treatment has finished
Priority
Warning
Modes
NIV
Possible Cause
NIV mode has finished
Clearway 2 action
Message will disappear after 5 seconds.
On Screen Message
“NIV Treatment has ended”
Power Fail
Item
Description
Definition
All power sources have been lost.
Priority
Medium
Modes
All
Possible Cause
All power sources have been lost.
Mains not connected, and batteries discharged
Clearway 2 action
Treatment not possible. Constant audio alarm
and flashing LED
On Screen Message
N/A