Host Commands Polycold Cryochiller
Circuit Sensor Readings Commands Installation and Operation Manual
Brooks Automation
7-38
214072 Revision B
CIRC_INLT_TEMP Circuit Inlet Temperature
CIRC_OTLT_TEMP Circuit Outlet Temperature
Parameter(s): [n],[valve],?
[n] is circuit number (1 or 2)
[valve] is Valve identifier:
0 = Cool
1 = Not Used
2 = Defrost
3 = Not Used
4 = Not Used
5 = Not Used
Return Value: [A][n] [n] = 0 valve closed [n] = 1 valve open
Default Value: N/A
Security Level Get: 1 Set: N/A
Example: “CIRCUIT_VALVE_STATE,1,0,?” Get circuit 1 cool valve state
Interface RS-232, Ethernet
Purpose: Read the circuit inlet temperature sensor value. TC9/TC11
Parameter(s): [n],?
[n] is circuit number (1 or 2)
Return Value: [A][nnn.n]
Default Value: N/A
Security Level Get: 1 Set: N/A
Example: “CIRC_INLT_TEMP,1,?” Get circuit 1 inlet temperature
Interface RS-232, Ethernet
Purpose: Read the circuit outlet temperature sensor value. TC10 / TC11
Parameter(s): [n],?
[n] is circuit number (1 or 2)
Return Value: [A][nnn.n]
Default Value: N/A
Security Level Get: 1 Set: N/A
Example: “CIRC_OTLT_TEMP,1,?” Get circuit 1 outlet temperature