Chapter 6 – Set-up and Display (Results) Messages
Measurement Set-up Commands
6.3.2 Transducer Sensitivity
This command allows you to enter the transducer sensitivity as shown on its cali-
bration chart. The syntax appears as shown in Fig.6.4.
The following rules apply to the specification of the sensitivity value:
Range: 0.0001E–12 to 10.0E–09
Resolution: 3 digits
Round: Nearest value represented by 3 digits (1.234E-10 is rounded to
1.23E-10)
Default: 1.0E–12
The syntax for the query, and the analyzer’s response, are shown in Fig.6.5.
Fig.6.4 The syntax for the Transducer_Sensitivity command
Fig.6.5 The syntax for the query Transducer_Sensitivity? and amplifier response