Measuring Ampifier Type 2525
User Manual Vol.2
6–5
BE1394 11
Chapter 6 Set-up and Display (Results) Messages
Measurement Set-up Commands
6.3.2 Transducer Sensitivity
This command allows you to enter the transducer sensitivity as shown on its cali-
bration chart. The syntax appears as shown in Fig.6.4.
The following rules apply to the specification of the sensitivity value:
Range: 0.0001E–12 to 10.0E–09
Resolution: 3 digits
Round: Nearest value represented by 3 digits (1.234E-10 is rounded to
1.23E-10)
Default: 1.0E–12
The syntax for the query, and the analyzers response, are shown in Fig.6.5.
Fig.6.4 The syntax for the Transducer_Sensitivity command
Fig.6.5 The syntax for the query Transducer_Sensitivity? and amplifier response
941147e
Te
CHANNEL_1 TRANSDUCER_SENSITIVITY
To 2525
:
SP
:
Sensitivity
NR
941148e
Te
CHANNEL_1
CHANNEL_1
TRANSDUCER_SENSITIVITY?
To 2525
:
Header OFf
Te
From 2525
: :
Value
NR
SP
:
TRANSDUCER_SENSITIVITY