Reference Values
18 / 96 DOC-M88-EXX157_V17_04.2019
Table3.4: Maximum allowed values for the line shape in terms of FWHM (Co-radiation). SC: scintillation
counter, SOL‑XE: energy dispersive solid state detector, position sensitive detector VÅNTEC‑1,
SSD160, LYNXEYE/ LYNXEYEXE/ LYNXEYEXE‑T, LYNXEYE-2, SSD160-2 and EIGER2R500K
Reflection
hkl
Bragg-Brentano Parallel Beam
(Göbel Mirror)
SC/SOL‑XE
[°]
VÅNTEC‑1
[°]
LYNXEYE/
LYNXEYEXE/
LYNXEYEXE‑T/
SSD160/
LYNXEYE‑2/
SSD160-2
EIGER2R500K
[°]
SC/ SOL‑XE/
LYNXEYE/
LYNXEYEXE/
LYNXEYEXE‑T/
SSD160/
LYNXEYE‑2/
SSD160-2
1)
EIGER2R500K
2)
[°]
0 1 2 0.06 0.10 0.06 0.22
1 0 4 0.06 0.10 0.06 0.22
3 0 0 0.10 0.11 0.10 0.22
2 2 6 0.18 0.18 0.18 0.27
1)
LYNXEYE / LYNXEYE XE / LYNXEYE XE-T / SSD160/ LYNXEYE‑2/ SSD160‑2 in 0D
mode
2)
EIGER2R500K in 0D mode with 14mm*16mm opening
For Parallel Beam Geometry the 0.2° equatorial Soller is used as standard and optimum set
up. If different equatorial Sollers are used, FWHM values will increase accordingly.