E. Customer Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 65 / 96
Part II: Angular Accuracy, Instrument Response and Resolution (D8Series)
Anode: Cu Bragg-Brentano Geometry
Angular Accuracy: Difference Reference Position corr Instrument Response: I
rel
Instrument
Resolution
Maximum allowed 2θ deviation depends on height Maximum allowed intensity deviation for point detectors ± 10% FWHM
obs
< FWHM
max
2θ
obs
[°]
Differ-
ence
2θ
obs
Reference
Position
corr
Difference
Reference
Position
corr
2θ
exp
1)
[°]
I
obs
I
min
I
exp
1)
I
max
FWHM
obs
[°]
FWHM
max
2)
[°]
25.575 21.49 23.88 26.27
35.148 90.00 100.00 110.00
88.989 11.06 12.29 13.52
127.670 13.37 14.85 16.34
Reference Position corr
½(Difference
max
+Difference
min
)
Detector
3)
Current Reference Position Shift
2-Theta
3)
Corrected Reference Position
(Current Reference Position - Reference
Position corr)
Height Maximum allowed shift 2θ for specified accuracy of:
N: scaling factor; For details cf. chapter
Reference Values [}15]
Vertical 150 0,02° ± 0,01°
obs: observed 214 0,03° ± 0,015°
exp: expected 258 0.04° ± 0,02°
corr: corrected Horizontal 214 0,02° ± 0,01°
rel: relative 258 0,02° ± 0,01°
1)
see table 3.1. For linear detector configurations intensities might be outside allowed deviations for point detectors due to reasons explained in chapter
Reference Values [}15];
2)
see table 3.3;
3)
mark applicable: detector in case of Theta/Theta and 2-Theta in case of Theta/2Theta;
Depending on specific measurement setups, individual reflections may not be measurable. Then measurements must be restricted to accessible reflections.