E. Customer Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 59 / 96
Part I: Instrument Details (D8 Series)
Instrument Configuration D8 ADVANCE
Instrument Details Set Values for Validation
Diffractome-
ter
D8 ADVANCE
1)
Tube Power: 40kV / 40mA (for FL tubes)
2)
Tube Power: 40kV / 30mA (for FF tubes)
2)
Vertical
goniometer
Theta/Theta
Theta/2Theta
Height:
150
Radius:
Primary radius
Secondary radius
....................[mm]
....................[mm]
Sample stage Standard
Rotating
FLIP-STICK Sample position 5
AUTO CHANGER Sample position A1
Compact Cradle/ Compact
Cradle
Plus
Compact UMC
Monochro-
matisation
Kβ-Filter
Secondary monochromator
Focusing monochromator
Discriminator Settings
(LYNXEYEXE/
LYNXEYEXE‑T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr. [V] Veto Thr. [V]
Slits Fixed divergence slit Divergence slit: 0.3° (approx. 0.6mm)
Fixed antiscatter slit Antiscatter slit: 0.3° (approx. 0.6mm)
Variable divergence slit Divergence slit: 0.3° (approx. 0.6mm)
Variable antiscatter slit Antiscatter slit: 0.3° (approx. 0.6mm)
Motorized air scatter screen Automatic mode
Primary axial Soller slit Primary axial Soller slit: 2.5°
Secondary axial Soller slit Secondary axial Soller slit: 2.5°
Receiving slit Receiving slit 0.1mm