F. Customer Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 69 / 96
Part I: Instrument Details (D8 Series)
Instrument Configuration D8 ADVANCE
Instrument Details Set Values for Validation
Diffractometer D8 ADVANCE
1)
Tube Power: 35kV / 40 mA (for FL tubes)
2)
Tube Power: 35kV / 30 mA (for FF tubes)
2)
Vertical
goniometer
Theta/Theta
Theta/2Theta
Height:
150
Radius:
Primary radius
Secondary radius
....................[mm]
....................[mm]
Sample stage Standard
Rotating
FLIP-STICK Sample position 5
AUTO CHANGER Sample position A1
Compact Cradle/ Compact
Cradle
Plus
Compact UMC
Monochroma-
tisation
Kβ-Filter
Secondary monochromator
Focusing monochromator
Discriminator Settings
(LYNXEYEXE/
LYNXEYEXE‑T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr. [V] Veto Thr. [V]
Slits Fixed divergence Antiscatter slit: 0.3° (approx. 0.6 mm)
Fixed antiscatter slit Antiscatter slit: 0.3° (approx. 0.6mm)
Variable divergence slit Divergence slit: 0.3° (approx. 0.6mm)
Variable antiscatter slit Antiscatter slit: 0.3° (approx. 0.6mm)
Motorized air scatter screen Automatic mode
Primary axial Soller slit Primary axial Soller slit: 2.5°
Secondary axial Soller slit Secondary axial Soller slit: 2.5°
Receiving slit Receiving slit 0.1mm