H. Customer Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 89 / 96
Instrument Configuration D8 DISCOVER
Instrument Details Set Values for Validation
Diffractome-
ter
D8 DISCOVER Tube Power: 35kV/40mA (for FL tubes)
Tube Power: 35kV/ 30mA (for FF tubes)
Vertical
goniometer
Theta/Theta
Theta/2Theta
Height:
150
214
258
Radius:
Primary radius
Secondary radius
....................
[mm]
....................
[mm]
Horizontal
goniometer
Theta/2Theta
Height:
214
258
Radius:
Primary radius
Secondary radius
....................
[mm]
....................
[mm]
Sample
stage
Standard
Rotating
FLIP-STICK Sample position 5
AUTO CHANGER Sample position A1
Compact Cradle/ Compact
Cradle
Plus
Centric Eulerian Cradle
UMC 150/151/1516/150 HTC
Compact UMC
Monochro-
matisation
Primary Göbel mirror 60 mm Exit slit 1.2mm
Discriminator Settings
(LYNXEYE XE/ LYNXEYE XE-T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr.
[V]
Veto Thr. [V]
Slits Primary axial Soller slit
Secondary equatorial Soller slit