EasyManuals Logo

Bruker D8 Series User Manual

Bruker D8 Series
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
H. Customer Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 89 / 96
Instrument Configuration D8 DISCOVER
Instrument Details Set Values for Validation
Diffractome-
ter
D8 DISCOVER Tube Power: 35kV/40mA (for FL tubes)
Tube Power: 35kV/ 30mA (for FF tubes)
Vertical
goniometer
Theta/Theta
Theta/2Theta
Height:
150
214
258
Radius:
Primary radius
Secondary radius
....................
[mm]
....................
[mm]
Horizontal
goniometer
Theta/2Theta
Height:
214
258
Radius:
Primary radius
Secondary radius
....................
[mm]
....................
[mm]
Sample
stage
Standard
Rotating
FLIP-STICK Sample position 5
AUTO CHANGER Sample position A1
Compact Cradle/ Compact
Cradle
Plus
Centric Eulerian Cradle
UMC 150/151/1516/150 HTC
Compact UMC
Monochro-
matisation
Primary Göbel mirror 60 mm Exit slit 1.2mm
Discriminator Settings
(LYNXEYE XE/ LYNXEYE XE-T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr.
[V]
Veto Thr. [V]
Slits Primary axial Soller slit
Secondary equatorial Soller slit

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker D8 Series and is the answer not in the manual?

Bruker D8 Series Specifications

General IconGeneral
ManufacturerBruker
SeriesD8
X-ray SourceSealed tube or rotating anode
GoniometerVarious types available, including θ/θ, θ/2θ, and vertical goniometers
Sample StageVarious options including XYZ stages and sample changers
SoftwareDIFFRAC.SUITE

Related product manuals