Care and Maintenance
88 ©2020 CAE 905K640152 v1.1
CAEAres
Step 1: Clean the Manikin
To clean the simulator, use a soft cloth and warm water with a mild detergent to wipe off any
moulage materials on the manikin’s skin. Gently rub the soiled area with a soft cloth to remove most
marks and stains. DO NOT use solvents or abrasive pads.
Inspect the manikin for damage or puncture marks. If damage exists, contact CAE Healthcare
Customer Service immediately for a repair. Prompt repair prevents expansion of the damaged area.
Step 2: Draining and Flushing the IV
If the simulation included use of optional IV administration equipment, perform the following
procedure to remove fluids prior to storing the manikin:
a. Drain the remaining fluid from the IV tubes in the left arm.
b. If simulated blood was used, flushing the tubes with clear distilled water is
recommended.
c. Fill a syringe with clear distilled water.
d. Attach the syringe to the appropriate IV prime port.
e. Infuse until clear fluid flows from the IV drain tubing.
Flushing the IV Fill/Drain Tubing
f. Disconnect the syringe and fill the syringe with air.
g. Flush the IV prime port with air until air flows from the IV drain tubing.