EasyManua.ls Logo

Campbell CR6 - Measuring Ground Reference Offset Voltage

Campbell CR6
285 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
For example, if 3 µV offset exists in the measurement circuitry, a 5 mV signal is measured as 5.003
mV. When the input or excitation is reversed, the second sub-measurement is –4.997 mV.
Subtracting the second sub-measurement from the first and then dividing by 2 cancels the offset:
5.003 mV (–4.997 mV) = 10.000 mV
10.000 mV / 2 = 5.000 mV
Ratiometric differential measurement instructions allow both RevDiff and RevEx to be set
True. This results in four measurement sequences, which the data logger processes into the
reported measurement:
l positive excitation polarity with positive differential input polarity
l negative excitation polarity with positive differential input polarity
l positive excitation polarity with negative differential input polarity
l negative excitation polarity with negative differential input polarity
For ratiometric single-ended measurements, such as a BrHalf(), setting RevEx = True
results in two measurements of opposite excitation polarity that are subtracted and divided by 2
for offset voltage reduction. For RevEx = False for ratiometric single-ended measurements,
an offset-voltage measurement is determined from self-calibration.
When the data logger reverses differential inputs or excitation polarity, it delays the same settling
time after the reversal as it does before the first sub-measurement. So, there are two delays per
measurement when either RevDiff or RevEx is used. If both RevDiff and RevEx are True,
four sub-measurements are performed; positive and negative excitations with the inputs one way
and positive and negative excitations with the inputs reversed. The automatic procedure then is
as follows:
1. Switch to the measurement terminals.
2. Set the excitation, settle, and then measure.
3. Reverse the excitation, settle, and then measure.
4. Reverse the excitation, reverse the input terminals, settle, measure.
5. Reverse the excitation, settle, measure.
There are four delays per measurement. In cases of excitation reversal, excitation time for each
polarity is exactly the same to ensure that ionic sensors do not polarize with repetitive
measurements.
Read More: The Benefits of Input Reversal and Excitation Reversal for Voltage Measurements.
9.10.8.2 Measuring ground reference offset voltage
Single-ended and differential measurements without input reversal use an offset voltage
measurement with the PGIA inputs grounded. This offset voltage is subtracted from the
9. Tips and troubleshooting 159

Table of Contents

Related product manuals