14
TE-BBR Instruction and Operation Manual
Exchanger/Refrigeration
Place the Refrigeration Unit (BBR model) or Air/Water Heat Exchanger (TE-BBR model) on the floor
below the Control Unit. Do not place the unit on the same table as the Control Unit as the fan/motor
vibration can interfere with the sensitive measuring system of the Load Unit. Also, it is necessary to
provide 15 cm (6") of clearance on all sides of the unit for sufficient air circulation. Clearance
requirements for refrigeration units may vary depending on the model.
Note: If a compressor has been purchased to supply compressed gas, place it on
the floor or at a location removed from the test apparatus to prevent the
compressor vibration from interfering with the Load Frame measurements.