Page 134
Error
Cause (A.) and Corrective Action (B.)
No IDS
A. This message is displayed to indicate that no ID
memory files have been created when trying to
access a memory location.
B. Refer to section 8.6 Using Memory/ID Storage to
create an ID memory file.
notArE
A. The NET/GROSS key was pressed when there
is no stored tare.
B. Perform either a push button tare operation or
using the numeric keypad enter a new tare
weight value.
oCAP
A. The load on the scale exceeds the capacity of
the scale.
B. Remove the excess load. Note that this could
indicate the scale may need recalibrated.
-oF-
A. The indicator is trying to display a positive
number with more than six digits or a negative
number with more than five digits.
B. Return to Gross weight mode and review the
Tare weight. Might indicate incorrect calibration.
UnStb
A. Motion is present when trying to power up, print,
zero the scale or perform a push button tare
function.
B. Wait for a stable indication before printing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cardinal 212 Series and is the answer not in the manual?

Cardinal 212 Series Specifications

General IconGeneral
BrandCardinal
Model212 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals