EasyManuals Logo

Cardinal 212 Series User Manual

Cardinal 212 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
212/212X Installation & Technical
8200-M639-O1 Rev D 09/14
Page 135
Error
Cause (A.) and Corrective Action (B.)
ConFig
A. Indicates improper stored calibration data, calibration
required.
B. Calibrate the scale with test weights. Consult your
scale service provider.
ErrA h
A. The load cell input is above the range of the indicator.
This may indicate a bad load cell or circuit failure.
B. Check for improper load cell wiring or excessive load.
Consult your scale service provider.
ErrA L
A. The load cell input is below the range of the indicator.
This may indicate a bad load cell or circuit failure.
B. Check for improper load cell wiring. Consult your scale
service provider.
Lo Bat
A. The (asterisk) on the lower left side of the 212
display or the Lo Bat annunciator on the 212X is used
to indicate the battery does not have a sufficient charge.
B. The battery should be charged or replaced.
Lo bat
A. This will be displayed just before the battery voltage
drops to a level where operation is affected. At this
level, the indicator will automatically turn OFF.
B. The battery will need to be charged or replaced before
operations can continue.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cardinal 212 Series and is the answer not in the manual?

Cardinal 212 Series Specifications

General IconGeneral
BrandCardinal
Model212 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals