212/212X Installation & Technical
8200-M639-O1 Rev D 09/14
Page 135
Error
Cause (A.) and Corrective Action (B.)
ConFig
A. Indicates improper stored calibration data, calibration
required.
B. Calibrate the scale with test weights. Consult your
scale service provider.
ErrA h
A. The load cell input is above the range of the indicator.
This may indicate a bad load cell or circuit failure.
B. Check for improper load cell wiring or excessive load.
Consult your scale service provider.
ErrA L
A. The load cell input is below the range of the indicator.
This may indicate a bad load cell or circuit failure.
B. Check for improper load cell wiring. Consult your scale
service provider.
Lo Bat
A. The (asterisk) on the lower left side of the 212
display or the Lo Bat annunciator on the 212X is used
to indicate the battery does not have a sufficient charge.
B. The battery should be charged or replaced.
Lo bat
A. This will be displayed just before the battery voltage
drops to a level where operation is affected. At this
level, the indicator will automatically turn OFF.
B. The battery will need to be charged or replaced before
operations can continue.