13. RETINAL FIELDS
The following fields can be acquired (see Figure 31 for fields representation for the right eye):
a. CENTRAL: centered on the foveal pit;
b. NASAL: centered 17° nasally to the foveal pit ;
c. TEMPORAL: centered 17° temporally to the foveal pit ;
d. SUPERO-TEMPORAL: centered 12° superiorly and 12° temporally to the foveal pit;
e. PERI-CENTRAL: centered 5° nasally to the foveal pit ;
f. SUPERIOR: centered 17° superiorly to the foveal pit ;
g. INFERIOR: centered 17° inferiorly to the foveal pit;
Figure 31 - DRS retinal fields
The field(s) acquired respectively when using the SINGLE FIELD or MULTI FIELD acquisition
modes described previously can be configured in the Settings (see par. 14).
a
optic disk
b
c
d
e
f
g