Remote Interface Description
5-5
| | | | :PART?
| | | | :LOT <lot number>
| | | | :LOT?
| | | | :SERIal <serial number>
| | | | :SERIal?
5.2.2 Command Description
IEEE 488.2 Command
*CLS
Clear status data configuration the following actions are needed.
Clear standard event status register
Clear status byte register except for MAV bit (bit 4).
*ESE < metric system value>
Use setting standard event status enables register value, <metric system value> range
is 0 255.
*ESE?
The controller is used for query standard event status of device enable register value.
The output format is <metric system value>, its’ range is 0 255.
*ESR?
The controller queries the standard event register value of the device. After performing
this command, the standard event register value will be cleared to 0.
The output format is <metric system value>, its’ range is 0 255.
*IDN?
The controller is for reading the basic data of the device. The output format separate four
fields by comma, it denote separately: manufacturer, device model, serial number,
firmware version.
*OPC
Operation complete command
*OPC?
Operation complete query command
The output format is ASCII character” 1 “.
*PSC 0 | 1
Power on status clear command
*PSC?
Power on status, clears query command. The output format is ASCII character” 1 “ or
” 0 “.
*RST
The device reset command.
*RCL < metric system value>