Ground Bond Tester 19572 Users Manual
5-6
Recall command
This command is recall the saved status.
*SAV < metric system value>
Save command
This command is for saving the current status to the memory.
*SRE < metric system value>
It is for setting service request register value, its’ <metric system value> value is 0 255.
*SRE?
The controller is for reading service request enable register initial setting. The output
format is <metric system value>, its’ range is 0 255.
*STB?
The controller is for reading status bit register value.
The output format is <metric system value>, its’ range is 0 255
SCPI Command
:MEMory:DELete:LOCAtion < register number >
This command is for deleting < register number > denoted parameter data in system
memory. < register number > is integral data.
:MEMory:FREE:STATe?
This command is for query remainder PRESET parameter quantity in the system memory.
:MEMory:FREE:STEP?
This command is for querying remainder STEP quantity in the system memory.
:MEMory:STATe:DEFine <name>, <register number>
This command is for defining a name to a memory located at a certain specified position
inside the main memory.
:MEMory:STATe:DEFine? <name>
This command is for querying the location inside the main memory by the memory name.
:MEMory:STATe:LABEl? < register number >
This command is for querying the name inside the main memory by the location inside the
main memory.
:MEMory:NSTates?
This command is for querying the main system *SAV / *RCL maximum value add 1 of can
be used parameter.
:SYSTem:ERRor[:NEXT]?
This command is for reading the message in error queue. Return message, please check
error messages in section 5.3.
:SYSTem:VERSion?
This command is for querying the system SCPI version.