Ground Bond Tester 19572 Users Manual
5-8
[:SOURce]:SAFEty:RESult:ALL:MMETerage?
This command is for querying MEASURE METER reading of all steps.
[:SOURce]:SAFEty:RESult:ALL:MODE?
This command is for querying MODE of all steps, it will return character data GB.
[:SOURce]:SAFEty:RESult:ALL:TIME[:ELAPsed][:TEST]?
This command is for querying test time of all steps.
[:SOURce]:SAFEty:RESult:ALL[:JUDGment]?
This command is for querying all judgment result. Return format: First Step Result,
Second Step Result, ..., Last Step Result. Code meaning as below table:
Test result code table:
Mode
GB
Code
HEX
DEC
STOP
70
112
USER STOP
71
113
CAN NOT TEST
72
114
TESTING
73
115
PASS
74
116
HIGH FAIL
11
17
LOW FAIL
12
18
OUTPUT A/D OVER
16
22
METER A/D OVER
17
23
[:SOURce]:SAFEty:RESult:COMPleted?
This command is for querying the system if complete execution action or not, return 1 or 0.
[:SOURce]:SAFEty:RESult[:LAST][:JUDGment]?
This command is for querying judgment result code of last STEP.
The judgment result code table:
Result
Code (hexadecimal)
Code (metric system)
PASS
74
116
USER STOP
71
113
CAN NOT TEST
72
114
TESTING
73
115
STOP
70
112
The judgment result fail code table:
GB MODE
Hex
Dec
HI
11
17
LO
12
18
ADV OVER
16
22
ADI OVER
17
23
[:SOURce]:SAFEty:RESult:LAST:MMETerage?
It queries Measure Meter value of the last step.
[:SOURce]:SAFEty:RESult:LAST:OMETerage?

Table of Contents