Ground Bond Tester 19572 Users Manual
5-10
[:SOURce]:SAFEty:STEP<n>:GB[:LEVel]?
This command is for querying selected STEP, ground resistance test needed current
value.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit[:HIGH] < numeric value >
This command is for setting selected STEP, ground resistance judgment high limit, the
unit is Ohm.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit[:HIGH]?
This command is for querying selected STEP, ground resistance judgment high limit.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit:LOW
This command is for setting selected STEP, ground resistance judgment low limit.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit:LOW?
This command is for querying selected STEP, ground resistance judgment low limit.
[:SOURce]:SAFEty:STEP<n>:GB:TIME[:TEST] < numeric value >
This command is for setting selected STEP, its’ test needed time; the unit is second (s).
[:SOURce]:SAFEty:STEP<n>:GB:TIME[:TEST]?
This command is for querying selected STEP, its’ test needed time.
[:SOURce]:SAFEty:PRESet:TIME:PASS < numeric value >
This command is for setting BUZZER beeps time when passes the test.
<numeric_value> is a value between 0.2 and 99.9(included).
[:SOURce]:SAFEty:PRESet:TIME:PASS?
This command is for querying BUZZER beeps output continuous time when passes the
test. It returns a value between 0.2 and 99.9(included).
[:SOURce]:SAFEty:PRESet:TIME:STEP < numeric value > | KEY
This command is for setting interval time between STEP, and the next time to start
command to execute the next STEP.
[:SOURce]:SAFEty:PRESet:TIME:STEP?
This is for querying interval setting between steps, return value is KEY or the value of unit
is second.
[:SOURce]:SAFEty:PRESet:TIME:JUDGment < numeric value >
This command is for setting a period of time, a certain of conditions do not detect.
[:SOURce]:SAFEty:PRESet:TIME:JUDGment?
This command is for querying the time of waiting detection.
[:SOURce]:SAFEty:PRESet:GB:FREQuency < numeric value >
This command is for setting ground impedance test output current frequency.
[:SOURce]:SAFEty:PRESet:GB:FREQuency?
This command is for querying ground impedance test output current frequency.
[:SOURce]:SAFEty:PRESet:GB:VOLTage < numeric value >
This command is for setting ground impedance test open voltage.