Remote Operation
: INTerval
FREQency
[: {CW | IMMediate}]
VOLTage
[: LEVel][: IMMediate][:AMPLitude]
: AC
: DC
: LIMit
: AC
: DC
: PLUS
: MINus
: RANGe
[SOURce :] CURRent : LIMit
Description : This command sets the rms current limit of the AC source for
software protection.
Query Syntax : [SOURce :] CURRent : LIMit?
Parameters : <NR2>, valid range: 0.00 ~ maximum current spec. of the
specific model. (unit: A)
Return Parameters : <NR2>
[SOURce :] CURRent : DELay
Description : This command sets the delay time for triggering over current
protection.
Query Syntax : [SOURce :] CURRent : DELay?
Parameters : <NR2>, valid range: 0.0 ~ 5.0 (unit: 0.5 second)
Return Parameters : <NR2>
[SOURce :] CURRent : INRUsh : STARt
Description : This command sets the start time of the inrush current
measurement.
Query Syntax : [SOURce :] CURRent : INRUsh : STARt?
Parameters : <NR2>, valid range: 0.0 ~ 999.9 (unit: ms)
Return Parameters : <NR2>
[SOURce :] CURRent : INRUsh : INTerval
Description : This command sets the measuring interval of the inrush current
measurement.
Query Syntax : [SOURce :] CURRent : INRUsh : INTerval?
Parameters : <NR2>, valid range: 0.0 ~ 999.9 (unit: ms)
Return Parameters : <NR2>
[SOURce :] FREQuency [: {CW | IMMediate}]
Description : The command sets the frequency of the output waveform of the
AC source in Hz.
Query Syntax : [SOURce :] FREQuency [: {CW | IMMediate}]?
Parameters : <NR2>, valid range: 15.00 ~ 1000.00 (unit: Hz)
Return Parameters : <NR2>
[SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : AC
Description : This command sets the AC composition of output voltage in
Volts.
Query Syntax : [SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : AC?