Programmable AC Source 61605 User’s Manual
Parameters : <NR2>, valid range: 0.0 ~ 150.0 (in low range), 0.0 ~ 300.0 (in
high range)
Return Parameters : <NR2>
[SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : DC
Description : This command sets the DC composition of output voltage in
Volts.
Query Syntax : [SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : DC?
Parameters : <NR2>, valid range: -212.1 ~ 212.1 (in low range), -424.2 ~
424.2 (in high range)
Return Parameters : <NR2>
[SOURce :] VOLTage : LIMit : AC
Description : This command sets the setting of Vac LIMIT which will restrict
the value of Vac.
Query Syntax : [SOURce :] VOLTage : LIMit : AC?
Parameters : <NR2>, valid range: 0.0 ~ 300.0 (unit: V)
Return Parameters : <NR2>
[SOURce :] VOLTage : LIMit : DC : PLUS
Description : This command sets the setting of Vdc LIMIT(+) which will
restrict the value of Vdc.
Query Syntax : [SOURce :] VOLTage : LIMit : DC : PLUS?
Parameters : <NR2>, valid range: 0.0 ~ 424.2 (unit: V)
Return Parameters : <NR2>
[SOURce :] VOLTage : LIMit : DC : MINus
Description : This command sets the setting of Vdc LIMIT(-) which will restrict
the value of Vdc.
Query Syntax : [SOURce :] VOLTage : LIMit : DC : MINus?
Parameters : <NR2>, valid range: 0.0 ~ -424.2 (unit: V)
Return Parameters : <NR2>
[SOURce :] VOLTage : RANGe
Description : This command sets output voltage range with three options of
LOW(150 V), HIGH(300 V), or AUTO.
Query Syntax : [SOURce :] VOLTage : RANGe?
Parameters : LOW | HIGH | AUTO
Return Parameters : LOW | HIGH | AUTO
7.6.2.4 CONFIGURE Sub-system
[SOURce :]
CONFigure
: INHibit
: EXTernal
: COUPling
[SOURce :] CONFigure : INHibit?
Description : This command sets REMOTE INHIBIT state. There are three
states for the feature of remote inhibit : OFF, LIVE, and TRIG.
Query Syntax : [SOURce :] CONFigure : INHibit?
Parameters : OFF | LIVE | TRIG