Remote Operation
8-17
8.6.2.6 CONFIGURE Sub-system
[SOURce :]
CONFigure
: INHibit
[SOURce :] CONFigure : INHibit?
Description : This command sets REMOTE INHIBIT state. There are three
states for the feature of remote inhibit: OFF, LIVE, and TRIG.
Query Syntax : [SOURce :] CONFigure : INHibit?
Parameters : OFF | LIVE | TRIG
Return Parameters : OFF | LIVE | TRIG
8.6.2.7 PHASE Sub-system
[SOURce :]
PHASe
: ON
: OFF
[SOURce :] PHASe : ON
Description : This command sets the transition angle of the waveform when it
out. Default DEGREE ON is 0 degree.
Query Syntax : [SOURce :] PHASe : ON?
Parameters : <NR2>, valid range: 0.0 ~ 359.9
Return Parameters : <NR2>
[SOURce :] PHASe : OFF
Description : This command sets the transition angle of the waveform when it
quit.
Query Syntax : [SOURce :] PHASe : OFF?
Parameters : <NR2>, valid range: 0.0 ~ 360.0, 360.0: mean is IMMED.
Return Parameters : <NR2>
8.6.2.8 TRACE Sub-system
TRACe
: RMS
TRACe
Description : This command sets waveform data of user-defined. It needs 1024
data points to construct a period of waveform.
Users have to normalize the data as the maximum point equal to
32767 or the minimum point equal to -32767.
Syntax : TRACe <waveform_name>, <amplitude> {,<amplitude>}
Parameters : <waveform_name>:US<n>, n=1~6, <amplitude>:<NR1>,valid range :
-32767 ~ 32767.
Example : TRACe US1,100,200,…,32767,...,500,800 <= 1024 points
This command needs about 5 sec to execute.