Programmable AC Source 61705 Users Manual
8-20
[SOURce :] LIST : FREQuency : STARt
Description : This command sets the sequence of start frequency list
points.
Query Syntax : [SOURce:] LIST : FREQuency : STARt?
Parameters : <NR2>, …, <NR2> valid range: 15.00 ~ 1200.00 (unit: Hz)
Return Parameters : <NR2>, …, <NR2>
[SOURce :] LIST : FREQuency : END
Description : This command sets the sequence of end frequency list points.
Query Syntax : [SOURce:] LIST : FREQuency : END?
Parameters : <NR2>, …, <NR2> valid range: 15.00 ~ 1200.00 (unit: Hz)
Return Parameters : <NR2>, …, <NR2>
[SOURce :] LIST : DEGRee
Description : This command sets the sequence of phase angle list points.
Query Syntax : [SOURce:] LIST : DEGRee?
Parameters : <NR2>, …, <NR2> valid range: 0.0 ~ 359.9
Return Parameters : <NR2>, …, <NR2>
OUTPut : MODE
Description : This command sets the operation mode.
Query Syntax : OUTPut : MODE?
Parameters : FIXED | LIST | PULSE | STEP | INTERHAR
Return Parameters : FIXED | LIST | PULSE | STEP | INTERHAR
TRIG
Description : This command sets LIST mode in OFF, ON execution state
after setting OUTPut : MODE LIST. If users want to change
the parameters, it’s necessary to set TRIG OFF then OUTPut :
MODE FIXED. Then, set OUTPut : MODE LIST again to get
ready to set TRIG ON.
Query Syntax : TRIG : STATE?
Parameters : OFF | ON
Return Parameters : OFF | RUNNING
8.6.2.10 PULSE Sub-system
[SOURce :]
PULSe
: VOLTage
: AC
: DC
: FREQuency
: SHAPe
: SPHase
: COUNt
: DCYCle
: PERiod
OUTPut
: MODE