Programmable AC Source 61705 User’s Manual
8-22
[SOURce :] PULSe : PERiod
Description : This command sets the period of the PULSE mode.
Query Syntax : [SOURce :] PULSe : PERiod?
Parameters : <NR2>, valid range: 0.1 ~ 99999999.9 (unit: ms)
Return Parameters : <NR2>
OUTPut : MODE
Description : This command sets the operation mode
Query Syntax : OUTPut : MODE?
Parameters : FIXED | LIST | PULSE | STEP | INTERHAR
Return Parameters : FIXED | LIST | PULSE | STEP | INTERHAR
TRIG
Description : This command sets PULSE mode in OFF execution state
after setting OUTPut : MODE PULSE. If users want to
change the parameters, it’s necessary to set TRIG OFF then
OUTPut : MODE FIXED. Then, set OUTPut : MODE PULSE
again to get ready to set TRIG ON.
Query Syntax : TRIG : STATE?
Parameters : OFF | ON
Return Parameters : OFF | RUNNING
8.6.2.11 STEP Sub-system
[SOURce :]
STEP
: VOLTage
: AC
: DC
: FREQuency
: SHAPe
: SPHase
: DVOLtage
: AC
: DC
: DFRequency
: DWELl
: COUNt
OUTPut
: MODE
TRIG
TRIG : STATE?
[SOURce :] STEP : VOLTage : AC
Description : This command sets the initial AC voltage of STEP mode.
Query Syntax : [SOURce :] STEP : VOLTage : AC?
Parameters : <NR2>, valid range: 0.0 ~ 150.0 (in low range), 0.0 ~ 300.0 (in
high range)
Return Parameters : <NR2>