Remote Operation
8-23
[SOURce :] STEP : VOLTage : DC
Description : This command sets the initial DC voltage of STEP mode.
Query Syntax : [SOURce :] STEP : VOLTage : DC?
Parameters : <NR2>, valid range: -212.1 ~ 212.1 (in low range), -424.2 ~
424.2 (in high range)
Return Parameters : <NR2>
[SOURce :] STEP : FREQuency
Description : This command sets the initial frequency of STEP mode.
Query Syntax : [SOURce :] STEP : FREQuency?
Parameters : <NR2>, valid range: 15.00 ~ 1200.00 (unit: Hz)
Return Parameters : <NR2>
[SOURce :] STEP : SHAPe
Description : This command selects the waveform buffer for STEP mode.
Query Syntax : [SOURce :] STEP : SHAPe?
Parameters : A | B
Return Parameters : A | B
[SOURce :] STEP : SPHase
Description : This command sets the start phase angle of STEP mode.
Query Syntax : [SOURce :] STEP : SPHase?
Parameters : <NR2>, valid range: 0.0 ~ 359.9
Return Parameters : <NR2>
[SOURce :] STEP : DVOLtage : AC
Description : This command sets the delta AC voltage in each step.
Query Syntax : [SOURce :] STEP : DVOLtage : AC?
Parameters : <NR2>, valid range: -150.0 ~ 150.0 (in low range), -300.0 ~
300.0 (in high range)
Return Parameters : <NR2>
[SOURce :] STEP : DVOLtage : DC
Description : This command sets the delta DC voltage in each step.
Query Syntax : [SOURce :] STEP : DVOLtage : DC?
Parameters : <NR2>, valid range: -212.1 ~ 212.1 (in low range), -424.2 ~
424.2 (in high range)
Return Parameters : <NR2>
[SOURce :] STEP : DFRequency
Description : This command sets the delta frequency in each step.
Query Syntax : [SOURce :] STEP : DFRequency?
Parameters : <NR2>, valid range: -1200.00 ~ 1200.00 (unit: Hz)
Return Parameters : <NR2>
[SOURce :] STEP : DWELl
Description : This command sets the dwell time in each step.
Query Syntax : [SOURce :] STEP : DWELl?
Parameters : <NR2>, valid range: 0.1 ~ 99999999.9 (unit: ms)
Return Parameters : <NR2>
[SOURce :] STEP : COUNt
Description : This command sets the number of times that the step is