Digital Power Meter 66200 Series User’s Manual
4. Full Scale Calibration for Current Measurement Range:
Low Current Shunt:
a. 0.01 Range Calibration:
(1) Connect the I+ and I- output of Fluke 5500A to the I+ and I- on 66202 rear
panel respectively. Then the screen will show as below (the V measurement
window blinks):
20
8
2
0.2
AUTO AUTOAUTO
150
300
500
VOLTAGE
CURRENT
SHUNT
2
0.4
0.1
0.01
LH
SYSTEM
Cali
Setup
Trig /
Enter
ON
OFF
I
O
GO/NG
PASS
Shunt
Limit
Meas
RMT
FAIL
Is/Trig
PF F
E
CF
THDv
THD
V
Vpk+
Vpk-
mA
I
Ipk+
Ipk-
Is
W
PF
VA
VAR
Chroma
DIGITAL POWER METER
MODEL
66202
k
(2) Press Trig/Enter the 10.00 in I measurement window blinks. Set the Fluke
5500A output to 150V/10mA/60Hz and then press OPR to begin output. (To
use other voltage for calibration, use , to adjust the value in I
measurement window and set the voltage in Fluke 5500A.)
(3) Press Trig/Enter the W measurement window will show the calibrated value
10.01 (this is the reference) and the PF window will show the message “done”.
The meter will beep concurrently.
20
8
2
0.2
AUTO AUTOAUTO
150
300
500
VOLTAGE
CURRENT
SHUNT
2
0.4
0.1
0.01
LH
SYSTEM
Cali
Setup
Trig /
Enter
ON
OFF
I
O
GO/NG
PASS
Shunt
Limit
Meas
RMT
FAIL
Is/Trig
PF F
E
CF
THDv
THD
V
Vpk+
Vpk-
mA
I
Ipk+
Ipk-
Is
W
PF
VA
VAR
Chroma
DIGITAL POWER METER
MODEL
66202
k
(4) Set the Fluke 5500A output to OFF and press Trig/Enter to go to next range
(0.1A) for calibration.
b. 0.1A Range Calibration:
(1) The calibration screen of 0.1A Range is shown as below (V measurement
window blinks):
F-8