Digital Power Meter 66200 Series User’s Manual
2. In the special measurement display ‘OFF’ (default) starts to blink. Use , key to set
it to ON and press Enter/Trig to save the setting and go to time setting for detection.
3. Use , key to select the time (second) required and press Enter/Trig to save the setting
and turn to parameter selection.
4. Select Vpk+ parameter in the AC voltage display. When the ‘V’ LED starts blinking
use , key to set it to Vpk+ and press Enter/Trig to save the setting and go to
High/Low limit setting.
5. Now the display has a blinking ‘H’. Press Enter/Trig will see the digits at right are
blinking. Use , to set the high limit and press Enter/Trig to save the setting and go
to Vpk+ LED.
6. Press Enter/Trig to enter the display again. Use , key to change the blinking ‘H’ to
‘L’ and press Enter/Trig will see the digits at right are blinking. Use , to set low
limit and Enter/Trig to save the setting and go to Vpk+ LED. The setting is now done.
7. To set other special parameters, please repeat step 4-6.
8. Press Setup to exit the function setting display.
When the above settings are done, the GO/NG LED will be on. Press Trig/Enter to start
measurement and specification assessment. If the parameters measured are within setting
range in test time, the 66200 Series Power Meter will beep twice when the test time is over
and the
PASS indicator will appear. On the contrary, the 66200 Series Power Meter will show
FAIL indicator and beep continuously if fail. The display will prompt the failed parameter
automatically and the test data will keep on blinking.
When PASS in on, press Trig/Enter will activate the next specification assessment. When
FAIL is on, it is necessary to press Trig/Enter to end the test for parameter reset or activate
the next specification assessment.
Follow the step below to disable the measurement of Limit parameter:
1. Press Setup to select Limit and the indicator of Limit is on.
2. In the special measurement display ‘ON’ starts to blink. Use , to set to ‘OFF’ and
press Enter/Trig to save the setting and exit the function setting.
GO/NG indicator will be off once it is disabled.
L NOTE
E (Energy), Is (Inrush current) and GO/NG three parameters of 66202 Power Meter need to
be triggered by Enter/Trig. Once two are set, the triggering priority is E > Is > GO/NG.
4.4.3 Meas Function
66200 Series Power Meter has five sub functions under Meas that are Meas, Energ, Power,
Flit and THD as explained below.
Meas:
66202 Power Meter provides two ways to show the panel readouts, which are Avg (average)
and Windo (window). The difference between two of them is the way of sampling as
described beneath.
4-10