Digital Power Meter 66200 Series User’s Manual
Reading + 0.33% of Range
PF accuracy 0.006+(0.003/PF)/KHz (*Note2)
0.027+(0.003/PF)/KHz
(*Note3)
Frequency
Range 15Hz – 10KHz 15Hz – 10KHz
Measuring Input Voltage (10% - 100% of the voltage
range)
Voltage (10% - 100% of the voltage
range)
Others
Display
Resolution
5 Digits
Display update
rate
0.5 sec
Power Supply 90V-130V /180V – 250V, 50Hz/60Hz, 30 VA
Interface Option: USB or GPIB+USB
Temperature
Coefficient Refer to temperature coefficient spec.
Operating 0 - 40°C
Storage -40 - 85°C
Safety & EMC
CE (include EMC & LVD)
Dimensions
(WxHxD)
212 x 88 x 348.1 mm (excluding projections)
Weight
3.8KG
Notes:
1. The current accuracy is permitted in 23°C±1°C for 0.01A /0.2A(CF=2) and in 23°C±5°C for the
other ranges.
2. The PF spec is permitted only if neither voltage nor current falls to be less than 50% of the range
rating.
3. The PF spec is permitted only if neither voltage nor current falls to be less than 10% of the range
rating.
2-2