Appendix F Calibration Procedure
Entering Calibration:
1. Power on the 66201.
2. Press Setup, Enter concurrently when the panel is in normal measurement state. The
screen appears as below:
20
8
2
0.2
AUTO AUTOAUTO
150
300
500
VOLTAGE
CURRENT
SHUNT
2
0.4
0.1
0.01
LH
SYSTEM
Cali
Setup
Trig /
Enter
ON
OFF
I
O
GO/NG
PASS
Shunt
Limit
Meas
RMT
FAIL
Is/Trig
PF F
E
CF
THDv
THD
V
Vpk+
Vpk-
mA
I
Ipk+
Ipk-
Is
W
PF
VA
VAR
Chroma
DIGITAL POWER METER
MODEL
66202
k
The screen shows P: 66201 F/W version
F: 66201 FPGA version
3. Press ↑, ↓ at the same time to enter into Calibration mode as shown below:
V
Vpk+
Vpk-
mA
I
Ipk+
Ipk-
Is
Use ↑, ↓ to select the screen display to YES (the YES screen blinks) and press Enter
to enter into to Calibration procedure.
To return to measurement mode without running calibration, press ↑, ↓ to select no
when YES is blinking, then press Trig/Enter to confirm will return to measurement mode.
F-3