K
p
K
i
K
d
Process
Variable
Error
Integral
Derivave
Output
85
4.6 Transducer Calibraon
Follow these steps to calibrate a transducer:
Idenfy the transducer output type.
Note: The ARControl can read Barksdale 422 series, mV/V, Raometric, 1-5V, and 4-20mA output type
transducers. The ARControl BMS provides non-incenve connecons for external Barksdale 422-series pressure
transducers. These transducers have been evaluated as non-incenve in this applicaon. They contain only a
resisve bridge without any capacitance or inductance.
1. Set the seng for SPAN:
a. Idenfy the output range of the transducer (e.g. -5 to 100 psi).
• Calculate the range (e.g. 100 psi -5 psi = 105 psi)
b. Determine the desired resoluon (e.g. 1 psi, 0.1 psi, 0.01 psi).
• Do not select a resoluon that is outside of the tolerance of the transducer
• The desired resoluon is the DECIMAL PLACE
c. Calculang XDCR
MAXOUTmV
(Note: Only applicable for mV/V transducers).
• Idenfy the maximum output of the transducer (located on label or datasheet)
• Convert this value to mV (millivolts) if some other unit e.g. 1.5V output = 1500mV; 10V = 10000mV
d. Ulizing the informaon from steps a through c, use the following equaons (Table 4.6.1) to calculate SPAN:
4.5 PID Control Loop
The system has a PID control loop which is ulized along with the accessory ARControl 4-20 mA module. The PID
controller allows for precise and stable closed-loop-control that is tolerant to system error. The PID controller requires
tuning which is best performed by personnel with PID tuning experience. The PID control loop runs at all mes, even in
DIAGNOSTIC mode.
Image 4.5.1 · PID control loop
Table 4.6.1 · SPAN equaons
SETTING DESCRIPTION
SPAN
XDCR Type SPAN Calculaon
BK 422
Span = (RANGE
xdcr
) ∙ 10
DECIMAL PLACE
mV/V
Span = (XDCR
MAXOUTmV
/ 156.25mV) ∙ (RANGE
xdcr
) ∙ 10
DECIMAL PLACE
Rao
Span = (RANGE
xdcr
) ∙ 10
DECIMAL PLACE
1-5V
Span = (RANGE
xdcr
) ∙ 10
DECIMAL PLACE
4-20mA
Span = (RANGE
xdcr
) ∙ 10
DECIMAL PLACE