Pulse oximeter charger
not working
Down level
software version
on pump
Release of new software
version
Upgrade software version of pump.
(ref. sec. 7)
Request for
specific
software version
Customer request Program pump with requested
software version
(ref. sec. 7)
Programmable
protocols are
still
programmed
Software default setting
have not been restored
Restore software default settings
(ref.
sec. 8)
If after
exhausting all
diagnostic and
service options
the pump still
does not operate
correctly.
Unknown Return pump to customer with
recommendation to remove from use.