Scheduled service table
Item
Number of
cycles or
max
allowed
time
Service date
Number of
cycles
performed at
intervention
operation
Technician signature
Air filter*
400
1 year
Water filter
1500
2 years
Vacuum pump kit
1000
2 years
EV2 drain
1500
2 years
EV6 by-pass
3000
3 years
Empty tank probe
1500
2 years
Water inlet pump
2000
2 years
Internal coil
6000
6 years
Door gasket*
3000
2 years
Chamber level probe
3000
3 years
Fan(s)
4000
3 years
Temperature probe 1
6000
6 years
Chamber safety thermostat
6000
6 years
Pressure transducer
6000
6 years
Chamber safety probe
6000
6 years
* this item can be replaced by the user.