EasyManua.ls Logo

COMINOX STERILCLAVE 18 - PT1;PT2;PT3 Probe Calibration; PT1;PT2;PT3 Correction

COMINOX STERILCLAVE 18
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
PT1 / PT2 / PT3 probe calibration
This procedure is used to complete the calibration of the temperature probes PT1, PT2 and
PT3. This operation must be performed whenever a temperature probe is replaced.
Press +/- to enter a 0°C value of the temperature probe shown on the label or on the wire
sheath (e.g.: if a value of 1000.2 is shown on the PT1 sensor label, press + twice to display
1000.2). Press ENTER to confirm.
Press to go back to the Technician Menu without changing the calibration.
PT1 / PT2 / PT3 correction
This procedure is used to correct the temperature value displayed during the sterilising
process. For example, if using an external sampling tool a temperature lower than the one
displayed is measured in the chamber during the sterilisation process, the temperature
may be corrected as follows:
Press +/- to enter the correction which will change the temperature shown on the display.
Press ENTER to confirm.
Press to go back to the Technician Menu without changing the correction.
Technician Menu

Table of Contents

Related product manuals