30
Cooling pause
This message indicates that the temperature in chamber 09 is too high to start or continue
a cycle (see USE AND MAINTENANCE MANUAL, chapter OPERATION).
The steriliser will display this message in STAND BY mode or after starting a cycle and will
stop only during the pre-vacuum phase. This message normally appears when several
cycles are run one after the other and no time is allowed for the chamber to cool down.
• In such conditions, simply let the chamber cool down. To speed up cooling, open
door 16, otherwise cooling will take considerably longer. To open the door, press STOP,
cancel the MANUAL STOP ALARM and wait for a few seconds for the pressure inside
the chamber to equalise ambient pressure to allow the door to be opened.
• With the door closed the operating cycle will resume automatically when the allowed
temperature is reached.
• If you opened the door, start the cycle again after cooling down the steriliser.
Messages