46
Vacuum pump servicing procedure
The vacuum pump can be serviced by cleaning it and replacing the seals which are
supplied in the specific kit.
Proceed as follows for vacuum pump 2119:
Remove the pump and carefully check the
position of components.
Check that the membranes are not cracked.
Clean or replace them.
Check the rubber valves. Clean or replaced
them. Make sure the position and orientation is
correct when refitting.
Vacuum pump kit 2119:
Vacuum pump 2119:
Additional Checks And Settings