EasyManua.ls Logo

COMINOX STERILCLAVE 18 - Additional Checks and Settings; Display off or Not Clear; Level Probe Check

COMINOX STERILCLAVE 18
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
ADDITIONAL CHECKS AND SETTINGS
Display off or not clear
Proceed as follows if no writing appears on the display and/or any if the display is not very
clear:
Check that power socket is powered, the power cable 03 is correctly plugged in
and connected to the steriliser.
Check fuses 32. Check that the main switch 13 is pressed to position “I”.
Check the fuses on electronic power board 34.
If the display is not clear from where you are standing, change reading position or
angle. If the display is clearer, adjust contrast by means of trimmer PT1 starting from
your preferred reading position (see chapter DIAGRAMS AND WIRING, Logic
electronic board).
Level probe check
Level probes 43 of both tanks are simple metal rods fitted into the container to be
monitored and insulated from it. The probe circuit is closed when liquid is present.
The presence of fouling or lime scales may create electric continuity between tank metal
and level probe rod and consequently read the presence of liquid in the tank when the
tank is empty.
Additional Checks And Settings

Table of Contents

Related product manuals