Door adjustment
• Procedure for 18/24 litre models: if leakages still occur after checking and cleaning
chamber seal 31 and the flange of chamber 09 on which it rests, you will need to:
- Switch the steriliser off and wait for it to cool down, then open the door.
- Use a wrench ch5 to loosen the screw inside the hole on the left side of the
door. When the screw is loose, the plate will be free to rotate.
- Turn the plate anticlockwise for about ¼ of a turn or as required.
- Tighten the Allen screw ch5.
- When closing the door, a higher resistance should be felt.
- Test tightness by starting a new cycle. Repeat the operation if required.
• For 6 S Dynamica, in case of leaks do the following:
- Check and clean chamber seal 31.
- Adjust the door closing by following these steps:
o Switch the steriliser off and wait for it to cool down, then open the
door.
o Remove the cover from the steriliser.
o Rotate the screw on top of the chamber to adjust the hook and thus
adjusting the door closing.
o Test tightness by starting a new cycle. Repeat the operation if
required.
Note: Do not tighten the door excessively. Excessive pressure may deform the seal.