Fit the automatic water inlet pump 46 correctly blocking it on its support.
Fit the branches (Y) and all connecting pipes.
Fit the external filling and drain pipes; the end of the inlet pipe is cut in a particular
manner to prevent a suction effect when it rests on the bottom of the tank. The
drain pipe is provided with a quick coupling male attachment.
Test tightness.
In the Technician Menu, Accessories option (option 16), set External Charge ON;
check operation of the steriliser. Close and restart the unit.
Automatic filling