Section 11: Utilities
158
To set the virtual DIP SW, begin with the DPU-414-30B switched off
and complete the following procedures:
• Slide the power switch to ON while pressing the ON LINE
button.
• Release the ON LINE button after a list of the current
settings starts printing out.
The printout of the current settings is followed by the prompt:
Continue ? :Press On-line SW
Write ? :Press Paper feed SW
• To change the DIP SW settings press the ON LINE button.
• To leave the DIP SW settings unchanged press the FEED
button.
After pressing the ON LINE button, DIP SW1 is printed, prompting
the input of new settings for switch numbers 1 through 8 of DIP SW1.
Refer to the following table for the required settings to print from the
instrument. Always input either ON or OFF for every setting for
switch numbers 1 through 8 because DIP SW Set mode cannot be
cancelled once it is initiated.
• ON can be set by pressing the ON LINE button once and
OFF by pressing the FEED button once
The setting is printed out after the ON LINE or FEED button is
pressed to confirm the new setting.
As soon as switch number 8 is set the printer once again prompts
Continue ? or Write ?. Continue ? (ON LINE) allows DIP SW2 and
DIP SW3 to be set in turn, and Write ? (FEED) exits DIP SW setting
mode.
After the FEED button is pressed, the ON LINE and OFF LINE lamps
alternately blink and the new settings are written to memory. When
the printer finishes writing everything to memory, 'DIP SW setting
complete !!' is printed out and the printer returns to the ON LINE
mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com