Section 3: Onsite Analysis - Measurement Menu
53
Default Parameters
When using the automated menus, the following parameters are
automatically applied.
Tach
Display
Time
Synchronous
Averaging
Bump
Test
Coast-
down/
Run-up
Cross
Channel
Phase
Domain Frequency Time Frequency Frequency Frequency
Window Hanning N/A Hanning Hanning Hanning
Spectral
Lines
400 User defined 400 400 400
Fmin 1 Hz N/A 1 Hz 1 Hz 1 Hz
Average
Type
4 X Linear 16 X Linear Peak hold 1 X Linear 1 X Linear
Amplitude
Scale
Linear N/A Linear Linear Linear
Trigger Tach
Free run
Tach
Free run
Internal
Free run
Internal
Free run
Tach
Free run
Display
Orders
OFF OFF OFF OFF OFF
Harmonic
Cursor
OFF OFF OFF OFF OFF
Tach
Display
ON OFF OFF OFF ON
Autorepeat OFF OFF OFF ON OFF
As with any measurement, the parameters can be changed before
and during measuring as described in Setting Measurement
Parameters (page
63).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Commtest Instruments vb Series and is the answer not in the manual?

Commtest Instruments vb Series Specifications

General IconGeneral
BrandCommtest Instruments
Modelvb Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals