6.2.7 Arc Parameters
ARC Parameters is used to enable/disable the ARC function and set the
ARC parameters. These settings are only to be changed by CPS
personnel with approval from the AHJ.
Figure 6-19 Arc Parameters Setting
6.2.8 Other Parameters
Other Parameters is used to set additional parameters including MPPT
scan period, nominal derating step and GFCI, DCI parameters. Press
ENT and use UP/DOWN keys to set parameters and enable/disable the
functions. Press ENT to confirm the setting. The parameters shown in
Figure 6-20 and Table 6-11.