EasyManua.ls Logo

CS Instruments VA 500 - Setting of Zeropoint and Low-Flow Cut off

CS Instruments VA 500
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
^ Operation
VA 500 Bi-directional English 1.24 Seite 36 von 56
11.3.1.5 Setting of Zeropoint and Low-flow cut off
Setup Sensor Setup ZP Adjust
1,03
Setup Sensor Setup ZP Adjust ZeroPnt
Setup Sensor Setup ZP Adjust CutOff
Setup Sensor Setup ZP Adjust t Reset
1,03
To make changes, first select a menu with
button and confirm selection by pressing
„OK“ .
When, without flow, the installed sensor shows
already a flow value of > 0 m³/h herewith the
zero point of the characteristic could be reset.
For an input / change of the value select with
the button the respective number position
and activate it with „OK“.
By pressing the position value is
incremented by 1. Confirm the input with „OK“
and activate next number position.
Leave menu with button „Back“
Menüpunkt mit Taste anhlen und
anschließend mit Taste OK“ auswählen
With the low-flow cut off activated, the flow
below the defined "LowFlow Cut off" value will
be displayed as 0 m³/h and not added to the
consumption counter.
For an input / change of the value select with
the button the respective number position
and activate it with „OK“.
By pressing the position value is
incremented by 1. Confirm the input with „OK“
and activate next number position.
Leave menu with button „Back“
By selection of „Reset“ all settings for
„ZeroPnt“ and. CutOff are reset.
Menu item to be select with button and
confirm the reset with „OK“ .
Leave menu with button „Back“

Table of Contents

Other manuals for CS Instruments VA 500

Related product manuals