EasyManua.ls Logo

Cygnus M3-DIVE - Two Point Calibration Procedure

Cygnus M3-DIVE
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cygnus DIVE Operating Manual
M3-DIVE-M-ENG_Iss9.docx
Page 58 of 117
Two Point Calibration Procedure
Two Point Calibration Procedure
1
First perform a Probe Zero function.
Probe Zero Screen
2
You must first determine the two
calibration thicknesses you will be
using. Then set these two thickness
values in the gauge’s Calibrate 2
Point menu.
Calibrate 2 Point menu
3
Set the minimum thickness value;
Select SET MIN THICKNESS from the
menu
Use the left and right buttons to
change the value press both to
save and exit.
Set Min Thickness
4
Set the maximum thickness value
Select SET MAX THICKNESS from the
menu
Use the left and right buttons to
change the value press both to
save and exit.
Set Max Thickness

Table of Contents

Related product manuals