Elrepho 3000
different measurement condition setup which is calibrated or
calibrate the desired measurement condition.
MODE NOT AVAILABLE UNTIL A ...(condition description)...
CONDITION CALIBRATED
Displayed whenever a measurement is requested using a
measurement condition which is not calibrated or not appropriate
for the operation being performed. Calibrate the measurement
condition specified in the message.
DIAGNOSTIC MEAS COND NOT CALIBRATED
Displayed while in any of the diagnostic test modes whenever a
measurement is requested using a measurement condition which is
not calibrated. Either select a different measurement condition
setup which is calibrated or calibrate the desired measurement
condition.
JOB NOT CALIBRATED
Displayed when asked to enter the Measure Mode and the current
job is not calibrated. This means that all the measurements required
for the job’s procedure can not be performed until more
measurement conditions are calibrated. The Calibration Mode
provides an option to calibrate the current job. Perform calibration
and then try again.
REMEASURE BLACK TRAP - UNSTABLE READING
This warning will be displayed during the black trap calibration if
there is too much variation in the reflectance of the first and second
measurements.
REMEASURE BLACK TRAP - TOO LIGHT
This warning will be displayed during the black trap calibration if
the measured black trap is too light. This usually indicates that
something other than the black trap is in position and was
measured. Check to ensure the black trap is positioned correctly.
REMEASURE WHITE TILE - UNSTABLE READING
This warning will be displayed during the white standard calibration
if there is too much variation in the reflectance of the first and
second measurements.
54 Datacolor International