UNABLE TO DETERMINE UVCAL STEP
This message indicates the UVCAL step position could not be
determined by performing measurements on the Illuminant
Checker. The minimum step number for the UVCAL position is
four. The maximum step number depends on the motor used to
control the UV filter position. Check that the Illuminant Checker’s
CIE Whiteness value is correct and that its surface is clean. If this
error still persists, this may indicate a possible instrument
malfunction and your Datacolor International Service representative
should be contacted for further details.
MUST SAVE AT LEAST ONE STD
This message will be displayed if you attempt to delete all white
calibration standards from memory. The system requires that at
least one be enabled so that calibration can be performed.
WHITENESS LESS THAN PREVIOUS STANDARDS
This message will be displayed when performing the Ganz/Griesser
Parameter Calibration and the nominal whiteness value entered for
the standard is less than the previous standard’s whiteness value.
The standards must be measured in increasing order.
WHITENESS VALUE MUST BE GREATER THAN ZERO
This message will be displayed when performing the Ganz/Griesser
Parameter Calibration and the nominal whiteness value entered for
the standard is set to zero when the Measure Button is pressed. The
standard’s must all have a whiteness value greater than zero.
RECAL INTERVAL INVALID
This message will be displayed if the recalibration time interval is
set to an invalid value. The interval must be set to some value
between zero and 999. Modify the setting to a value in the
acceptable range.